Test Code LAB10134 Amikacin - Trough
Department
CHEMISTRY
Collection Requirements
PEAK AND TROUGH DRUG LEVELS SHOULD BE DRAWN DURING THE SAME DOSE CYCLE WITH THE PEAK DRAWN (45-60 min post IM; immediately after 60 min. IV dose or 30 min after 30 min IV dose), AND THE TROUGH DRAWN (immediately prior to the next dose).
Primary Collection Container
Gold (SST)
Alternate Collection Container
Yellow Microtainer SST
Transport
Serum refrigerated
Preferred Transport Temperature
Refrigerated
Processing
Centrifuge and remove from gel if testing is delaed longer than 8 hours.
Stability
Refrigerated: 48 hours Frozen: 4 weeks
Minimum Testing Volume
0.5 mL
Reference Range
4.0-8.0 µg/mL
Methodology
Roche c503 homogenous immunoassay; Kinetic interaction of microparticles (KIMS) as measured by changes in light transmission
Performed
Sunday – Saturday
Reported
As completed
CPT Codes
80150
Unacceptable Conditions
Specimens stored longer than 8 hours on gel are unacceptable.