Test Code LAB10278 Glomerular Basement Membrane Antibodies, IgG, Serum

Reporting Name

Glomerular Basement Membrane IgG Ab

Useful For

Evaluating patients with clinical features of anti-glomerular basement disease, including rapidly progressive glomerulonephritis or pulmonary hemorrhage

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Ordering Guidance


If patient is being evaluated for autoimmune skin disease, order CIFS / Cutaneous Immunofluorescence Antibodies (IgG), Serum for evaluation of anti-intercellular substance (ICS) and antibasement membrane zone (BMZ) antibodies.



Specimen Required


Supplies: Sarstedt Aliquot Tube 5 mL (T914)

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 0.5 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.35 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 21 days
  Frozen  21 days

Reference Values

<1.0 U (negative)

≥1.0 U (positive)

Reference values apply to all ages.

Day(s) Performed

Monday through Friday

Test Classification

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

83516

LOINC Code Information

Test ID Test Order Name Order LOINC Value
GBM Glomerular Basement Membrane IgG Ab 31254-6

 

Result ID Test Result Name Result LOINC Value
GBM Glomerular Basement Membrane IgG Ab 31254-6

Clinical Information

Anti-glomerular basement (GBM) disease is a rare autoimmune disease, with an estimated incidence of 0.6-1.79 cases per million population per year.(1) Without prompt treatment, this disease is potentially fatal. Patients may present with rapidly progressive glomerulonephritis, pulmonary hemorrhage, or both.(2,3) The serological hallmark of this disease is the presence of anti-GBM antibodies of the IgG isotype. Anti-GBM antibodies bind to the non-collagenous domain 1 (NC1) of the alpha3 chain of type IV collagen, which is one of the main components of the kidney and lung basement membranes. Deposition of anti-GBM antibodies in the kidney and lungs triggers complement activation and production of reactive oxygen species, ultimately leading to vascular necrosis and damage to the GBM.

 

The diagnosis of anti-GBM disease in a patient with compatible clinical symptoms is often confirmed by detecting the presence of anti-GBM antibodies. This can be accomplished by a variety of antigen-specific, solid-phase immunoassays. Given the implications of this testing, understanding the diagnostic sensitivity and specificity of anti-GBM antibody methods is critical. In a recent meta-analysis, a pooled sensitivity of 93% (95%CI: 84-97%) and a pooled specificity of 97% (95%CI: 94-99%) was demonstrated across 11 methods.(4) In addition, some studies have suggested a prognostic role for anti-GBM antibodies, with higher titers being associated with increased mortality. However, it appears that this effect can largely be abrogated by prompt and aggressive treatment, particularly plasmapheresis.(1)

Interpretation

A positive result for anti-glomerular basement (GBM) antibody is consistent with the diagnosis of anti-GBM disease, in patients with the appropriate clinical presentation.

Method Description

Glomerular basement membrane (GBM) antigen is covalently coupled to polystyrene microspheres that are impregnated with fluorescent dyes to create a unique fluorescent signature. GBM antibodies, if present in diluted serum, bind to the GBM antigen on the microspheres. The microspheres are washed to remove extraneous serum proteins. Phycoerythrin (PE)-conjugated antihuman-IgG antibody is then added to detect IgG anti-GBM bound to the microspheres. The microspheres are washed to remove unbound conjugate, and bound conjugate is detected by laser photometry. A primary laser reveals the fluorescent signature of each microsphere to distinguish it from microspheres that are labeled with other antigens. A secondary laser reveals the level of PE fluorescence associated with each microsphere. Results are calculated by comparing the median fluorescence response for GBM microspheres to a 4-point calibration curve.(Package insert: Bio-Plex 2200 Vasculitis. Bio-Rad Laboratories; 12/2018)

Report Available

2 to 3 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus OK

Method Name

Multiplex Flow Immunoassay

Secondary ID

8106

Forms

If not ordering electronically, complete, print, and send a Renal Diagnostics Test Request (T830) with the specimen.