Sign in →

Test Code LAB12757 Platelet Inhibition (Plavix)

Department

HEMATOLOGY

Collection Requirements

Coveanant HealthCare - DO NOT TRANSPORT VIA TUBE SYSTEM

Draw a discard tube first.
Must be a full draw.
If a CBC is also ordered, collect the CBC last.
If drawing from a catheter, you must draw at least 5.0 mL of waste prior to drawing the Plavix specimen.

Primary Collection Container

2 Blue 2.0 mL NaCITRATE (Blue White)

Transport

Whole blood ambient

Preferred Transport Temperature

Ambient

Stability

Ambient: 4 hours

Refrigerated: Unacceptable

Frozen: Unacceptable

Minimum Testing Volume

2.0 mL

Reference Range

182-335 PRU

Values < 180 PRU suggest a drug effect from a P2Y12 inhibitor on platelet reactivity.

Methodology

VerifyNow P2Y12 Assay

Performed

Sunday – Saturday

Reported

As completed

CPT Codes

85576

Unacceptable Conditions

Hemolysis, clotted specimens, refrigerated or frozen specimens, specimens over 4 hours old.

Additional Information

Can also be used to verify if patient is therapeutic for Effient and Brilinta.

Patients with platelet values less than 119k or more than 502k and Hematocrit values less than 33% or greater than 52% can interfere with testing and valid results may not be obtained.

The P2Y12 Platelet Function assay is not intended for use with inherited platelet disorders such as von Willebrand Factor deficiency, Glanzmann Thrombasthenia and Bernard-Soulier Syndrome.