Test Code LAB20260 Atypical Hemolytic Uremic Syndrome Complement Panel, Serum and Plasma

Secondary ID

64881

Useful For

Detecting deficiencies in the alternative pathway that can cause atypical-hemolytic uremic syndrome, dense deposit disease, and C3 glomerulonephritis

 

A second-tier test that aids in the differential diagnosis of thrombotic microangiopathies

Profile Information

Test ID Reporting Name Available Separately Always Performed
INTGA AHUS Interpretation No Yes
COM3 Complement, Total, S Yes, (order COM) Yes
AH503 Alternative Complement Path Func, S Yes, (order AH50) Yes
C3HUS Complement C3, S Yes, (order C3) Yes
C4HUS Complement C4, S Yes, (order C4) Yes
FBCA Factor B Complement Antigen, S No Yes
FHCA Factor H Complement Antigen, S No Yes
C4D C4d Complement, P No Yes
CBB CBb Complement, P No Yes
SC5B9 SC5b-9 Complement, P Yes, (C5B9) Yes

Reflex Tests

Test ID Reporting Name Available Separately Always Performed
C1Q Complement C1q, S Yes No
C1QFX C1Q Complement, Functional, S Yes No
C2FXN C2 Complement, Functional, S, NR Yes No
C3FX C3 Complement, Functional, S Yes No
C4FX C4 Complement, Functional, S Yes No
C5FX C5 Complement, Functional, S Yes No
C6FX C6 Complement, Functional, S Yes No
C7FX C7 Complement, Functional, S Yes No
C8FX C8 Complement, Functional, S Yes No
C9FX C9 Complement, Functional, S Yes No
C5AG2 C5 Complement, Antigen, S Yes, (order C5AG) No

Method Name

C3HUS, C4HUS, FBCA, FHCA: Nephelometry

COM3: Automated Liposome Lysis Assay

AH503, C4D, CBB, SC5B9: Enzyme-Linked Immunosorbent Assay (ELISA)

Reporting Name

aHUS Complement Panel, S and P

Specimen Type

Plasma Na Cit
Serum Red


Ordering Guidance


This test should be performed prior to treatment initiation and in the absence of therapy with complement inhibitors, such as eculizumab or ravulizumab. Complement inhibitors will affect performance of these assays.

 

For evaluating patients with possible thrombotic microangiopathies (TMA), the recommended first-tier test is ADM13 / ADAMTS13 Activity and Inhibitor Profile, Plasma. This test should be a second-tier test for TMA.

 

For patients who have received eculizumab or need to monitor response to eculizumab therapy, the recommended test is ECUMP / Eculizumab Monitoring Panel, Serum. Soluble membrane attack complex (sMAC) should not be used as a standalone assay to monitor eculizumab efficiency.



Specimen Required


Both plasma and serum are required for this test.

 

Patient Preparation:

1. Fasting preferred.

2. Samples should not be collected earlier than 48 hours following plasma exchange.

 

Supplies: Sarstedt Aliquot Tube 5 mL (T914)

 

Specimen Type: Plasma

Collection Container/Tube: Light-blue top (3.2% sodium citrate)

Submission Container/Tube: 3 plastic vials

Specimen Volume: 1.5 mL in 3 plastic vials, each containing 0.5 mL

Collection Instructions:

1. Immediately after specimen collection, place the tube on wet ice.

2. Centrifuge; 1500 x g for 10 minutes at 4° C and aliquot plasma into plastic vial.

3. Freeze specimen within 30 minutes.

 

Specimen Type: Serum

Collection Container/Tube: Red top (serum gel/SST are not acceptable)

Submission Container/Tube: 3 plastic vials

Specimen Volume: 1.5 mL in 3 plastic vials, each containing 0.5 mL

Collection Instructions:

1. Immediately after specimen collection, place the tube on wet ice.

2. Centrifuge at 4° C and aliquot serum into 5 mL plastic vial.

3. Freeze specimen within 30 minutes.


Specimen Minimum Volume

Serum, Plasma: 1 mL each

Specimen Stability Information

Specimen Type Temperature Time Special Container
Plasma Na Cit Frozen 14 days
Serum Red Frozen 14 days

Reject Due To

Gross hemolysis OK
Gross lipemia Reject
Gross icterus OK

Clinical Information

Individuals presenting with thrombotic microangiopathies (TMA) require clinical testing to identify the underlying cause. Thrombotic thrombocytopenic purpura (TTP) and hemolytic uremic syndrome (HUS) are both acute syndromes with many overlapping clinical features. Reduced levels of ADAMTS13 (a disintegrin and metalloproteinase with thrombospondin type 1 motives, member 13) activity is associated with TTP and is one laboratory feature that distinguishes TTP from HUS. HUS can also have a number of causes; one of the rarer forms of disease is caused by defects in the alternative pathway of the complement system, so called atypical-HUS (aHUS). Patients with defective alternative pathway regulation can benefit from biologics that suppress the complement system.

 

The purpose of this panel is to aid in the differential diagnosis of TMA. The suggested approach is to rule-out other causes of TMA first, since aHUS is one of the rarer causes of TMA. Additionally, the assays can be used in the setting of membranoproliferative glomerulonephritis (MPGN) and can help distinguish between immune-complex mediated or complement-mediated kidney disease. MPGN mediated by immune-complexes are ones resulting from infectious processes, autoimmune diseases, or monoclonal gammopathies; whereas complement-mediated MPGN can be subdivided in C3 glomerulonephritis and dense deposit disease, based on electron microscopy of the kidney biopsy histological findings. Despite phenotypic differences, these glomerular diseases share dysfunction of the alternative pathway as the defining pathophysiology.

Reference Values

FACTOR B COMPLEMENT ANTIGEN

15.2-42.3 mg/dL

 

SC5b-9 COMPLEMENT

≤250 ng/mL

 

FACTOR H COMPLEMENT ANTIGEN

18.5 to 40.8 mg/dL

 

C4d COMPLEMENT ACTIVATION FRAGMENT

≤9.8 mcg/mL

 

CBb COMPLEMENT ACTIVATION FRAGMENT

≤1.6 mcg/mL

 

COMPLEMENT C4

14-40 mg/dL

 

COMPLEMENT C3

75-175 mg/dL

 

ALTERNATIVE COMPLEMENT, PATHWAY (AH50) FUNCTIONAL

≥46% normal

 

COMPLEMENT, TOTAL

30-75 U/mL

Interpretation

An interpretive report will be included.

Method Description

Complement, Total:

An automated method is performed using liposomes as the target for the serum complement system. The dinitrophenyl (DNP)-labeled liposomes are sensitized with antibody to DNP. Serum complement causes lysis and release of entrapped glucose-6-phosphate dehydrogenase. Glucose-6-phosphate dehydrogenase reacts with glucose-6-phosphate and nicotinamide adenine dinucleotide (NAD[+]). NAD(+) is reduced to NADH and the conversion is measured at 340 nm. The assay correlates with the CH50 assay based on sheep red blood cell lysis, has lower variability, and is simpler to perform.(Package insert: Fujifilm Autokit CH50. Fujifilm Wako Pure Chemical Corporation; 04/01/2018; Yamamoto S, Kubotsu K, Kida M, et al. Automated homogeneous liposome-based assay system for total complement activity. Clin Chem. 1995;41:586-590)

 

Alternative Complement, Pathway Functional:

The Wieslab enzyme-linked immunosorbent assay complement assay for the alternative pathway combines principles of the hemolytic assay for complement activation with the use of labeled antibodies specific for neoantigens produced as a result of complement activation. The micro titer plate strips are coated with lipopolysaccharide. Patient serum is diluted in diluent containing specific blocker to ensure that only the alternative pathway is activated. During the first incubation, the diluted patient serum in the wells is activated by the coating. The wells are then washed and C5b-9 (membrane attack complex: MAC) is detected with a specific alkaline phosphatase labeled antibody to the neoantigen expressed during MAC formation. After a final wash, an alkaline phosphatase substrate is added. The amount of alternative pathway complement activity correlates with the color intensity of the solution and is measured in terms of absorbance (optical density).(Frazer-Abel A, Sepiashvili L, Mbughuni MM, Willrich MA. Overview of laboratory testing and clinical presentations of complement deficiencies and dysregulation. Adv Clin Chem. 2016;77:1-75. doi:10.1016/bs.acc.2016.06.001)

 

Complements C3 and C4; Factor B and Factor H Complement Antigens:

In these Siemens Nephelometer II methods, the light scattered onto the antigen-antibody complexes is measured. The intensity of the measured scattered light is proportional to the amount of antigen-antibody complexes in the sample under certain conditions. If the antibody volume is kept constant, the signal behaves proportionally to the antigen volume.

 

A reference curve is generated by a standard with a known antigen content on which the scattered light signals of the samples can be evaluated and calculated as an antigen concentration. Antigen-antibody complexes are formed when a sample containing antigen and the corresponding antiserum are put into a cuvette. A light beam is generated with a light emitting diode, which is transmitted through the cuvette. The light is scattered onto the immuno-complexes that are present. Antigen and antibody are mixed in the initial measurement, but no complex is formed yet. An antigen-antibody complex is formed in the final measurement.

 

The result is calculated by subtracting value of the final measurement from the initial measurement. The distribution of intensity of the scattered light depends on the ratio of the particle size of the antigen-antibody complexes to the radiated wavelength.(Instruction manual: Siemens Nephelometer II Operations. Siemens, Inc; Version 2.3, 2008; Addendum to the Instruction Manual 2.3, 08/2017)

 

C4d Complement Activation Fragment:

Microtiter plates are coated with monoclonal antibody specific to the C4d fragment of the fourth component of the complement cascade. Controls, standards, and patient samples are exposed to the plate. After washing the plate, a horseradish peroxidase-conjugated polyclonal C4d antibody is added followed by a substrate to initiate color change.(Package insert: MicroVue C4d Fragment EIA Kit. Quidel Corporation; A009000EN00 10/2017)

 

CBb Complement Activation Fragment:

Microtiter plates are coated with monoclonal antibody specific to the complement factor Bb (CBb) fragment of the fourth component of the complement cascade. Controls, standards, and patient samples are exposed to the plate. After washing the plate, a horseradish peroxidase-conjugated polyclonal CBb antibody is added followed by a substrate to initiate color change.(Package insert: MicroVue CBb Plus EIA Kit. Quidel Corporation; 027002EN00 09/2016)

 

SC5b-9 Complement Activation Complex:

Microtiter plates are coated with monoclonal antibody specific to the C9 ring of the SC5b-9 complex. Controls, standards, and patient samples are exposed to the plate. After washing the plate, a horseradish peroxidase-conjugated anti-SC5b-9 complex antibody is added followed by a substrate to initiate color change.(Package insert: MicroVue SC5b-9 Plus EIA Kit. Quidel Corporation; 020001EN00 05/2017)

Day(s) Performed

Varies

Report Available

12 to 21 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

CPT Code Information

86160 x 7

86161

86162

LOINC Code Information

Test ID Test Order Name Order LOINC Value
AHUSD aHUS Complement Panel, S and P In Process

 

Result ID Test Result Name Result LOINC Value
62584 C4d Complement, P 39565-7
62585 CBb Complement, P 4517-9
FBCA Factor B Complement Antigen, S 2269-9
FHCA Factor H Complement Antigen, S 4519-5
62586 SC5b-9 Complement, P 93244-2
38316 Alternative Complement Path Func, S 74520-8
COM3 Complement, Total, S 4532-8
C3HUS Complement C3, S 4485-9
C4HUS Complement C4, S 4498-2
39844 AHUS Interpretation 69048-7
ECPRO Is Eculizumab or Ravulizumab taken? 86955-2

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

Forms

If not ordering electronically, complete, print, and send 1 of the following forms with the specimen:

-Renal Diagnostics Test Request (T830)

-Coagulation Test Request (T753)