Test Code LAB20971 TEG(Thrombelastograph) Platelet Mapping
Department
Hematology
Collection Requirements
Only available for Covenant HealthCare Hospital Inpatient Services- Venipuncture: A 21G or larger bore needle is required. In order to avoid tissue contaminants, draw a 2.7 mL blue top (sodium citrate) tube as a discard before drawing the tube for testing.
- Line draws: Draw from the appropriate port as per Covenant HealthCare policy and discard the first 7-10 mL. It is unacceptable to fill tubes from syringes when no discard syringe has been drawn as it will impact specimen quality and make the test results invalid.
- Draw one 2.7mL blue-top sodium citrate tube of blood to the appropriate level (excess anticoagulant may affect the test results). Use only 2.7mL sodium citrate tubes.
- Gently invert the tube 5 times to mix.
- Label the tube, verify specimen collection and place the specimen in a biohazard bag (with no additional specimens included) and immediately send the specimen over via the pneumatic tube to the Main Lab.
Primary Collection Container
Blue 2.7 ML (Citrate)
Green (Sodium Heparin)
Transport
Ambient
Preferred Transport Temperature
Ambient
Processing
Whole Blood
Stability
Ambient 2 Hours
Minimum Testing Volume
2.5 mL
Reference Range
HKH MA - 53-68 mmActF MA - 2-19 mm
ADP MA - 45-69 mm
AA MA - 51-71 mm
ADP % Inhibition - 0-17%
AA % Inhibition - 0-11%
Methodology
TEG 6s Hemostasis Syste
Performed
Sunday-Saturday
Reported
As completed
CPT Codes
85576, 85576 x3
Unacceptable Conditions
Frozen, refrigerated, clotted, hemolyzed, centrifuged or serum specimens.
Specimens greater than 2 hours old.