Sign in →

Test Code MISC2MAYOB210R BCR/ABL1, p210, mRNA Detection, Reverse Transcription-PCR (RT-PCR), Quantitative, Reflex, Varies


Specimen Required


Only orderable as a reflex. For more information see BCRFX /BCR/ABL1 Qualitative Diagnostic Assay with Reflex to BCR/ABL1 p190 Quantitative Assay or BCR/ABL1 p210 Quantitative Assay, Varies.

 

Submit only 1 of the following specimens:

 

Specimen Type: Whole blood

Container/Tube:

Preferred: Lavender top (EDTA)

Acceptable: Yellow top (ACD)

Specimen Volume: 10 mL

Collection Instructions:

1. Invert several times to mix blood.

2. Send whole blood specimen in original tube. Do not aliquot.

3. Label specimen as blood.

 

Specimen Type: Bone marrow

Container/Tube:

Preferred: Lavender top (EDTA)

Acceptable: Yellow top (ACD)

Specimen Volume: 4 mL

Collection Instructions:

1. Invert several times to mix bone marrow.

2. Send bone marrow specimen in original tube. Do not aliquot.

3. Label specimen as bone marrow.


Secondary ID

48390

Useful For

Diagnostic workup of patients with a high probability of BCR-ABL1-positive hematopoietic neoplasms, particularly chronic myeloid leukemia and Ph+ acute lymphoblastic leukemia (B-lymphoblastic leukemia), to provide a pretreatment quantitative level of BCR-ABL1 mRNA transcript if the initial diagnostic reverse transcription polymerase chain reaction screen is positive

 

When positive, the reflex test provides a quantitative value for the corresponding e13-a2 or e14-a2 (p210) BCR-ABL1 mRNA fusion variant

Method Name

Only orderable as a reflex. For more information see BCRFX /BCR/ABL1 Qualitative Diagnostic Assay with Reflex to BCR/ABL1 p190 Quantitative Assay or BCR/ABL1 p210 Quantitative Assay, Varies.

 

Quantitative Reverse Transcription-Polymerase Chain Reaction (RT-PCR)

Reporting Name

BCR/ABL1, p210, Quant, Reflex

Specimen Type

Varies

Specimen Minimum Volume

8 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Varies Refrigerated (preferred) 72 hours PURPLE OR PINK TOP/EDTA
  Ambient  72 hours PURPLE OR PINK TOP/EDTA

Reject Due To

Gross hemolysis Reject
Moderately to severely clotted Reject

Clinical Information

The t(9;22)/BCR-ABL1 abnormality is associated with chronic myeloid leukemia (CML) and "Philadelphia positive" acute lymphoblastic leukemia of B-cell lineage (Ph ALL). Very rarely, this abnormality has also been identified in cases of acute myeloid leukemia and T-lymphoblastic leukemia/lymphoma. The fusion gene on the derivative chromosome 22q11 produces a chimeric BCR-ABL1 messenger RNA (mRNA) transcript and corresponding translated oncoprotein. Despite substantial breakpoint heterogeneity at the DNA level, a consistent set of BCR-ABL1 mRNA transcripts are produced that can be readily and sensitively detected by reverse transcription polymerase chain reaction (RT-PCR) technique. In CML, breakpoints in BCR nearly always result in either exons 13 or 14 (e13, e14) joined to exon 2 of ABL1 (a2). The corresponding e13-a2 or e14-a2 BCR-ABL1 mRNAs produce a 210 kDa protein (p210). Rare cases of CML are characterized by an e19-a2 type mRNA with a corresponding p230 protein. In Ph ALL, the majority of cases harbor an e1-a2 BCR-ABL1 mRNA transcript, producing a p190 protein, although some ALL patients may alternatively present with the e13/e14-a2 or p210 type fusion.

 

This assay provides information at the time of diagnosis regarding the presence (and specific mRNA type) or absence of the BCR-ABL1 mRNA. If positive, the reflex test will follow to provide an initial quantitative level of the specific BCR-ABL1 transcript. For example, when positive for the e13/e14-a2 (p210) type mRNA, the reflex test provides a corresponding p210 quantitative value. Results from this test are also useful to determine the correct quantitative assay for subsequent monitoring of transcript levels (ie, p190 or p210) during tyrosine kinase inhibitor therapy.

Reference Values

Only orderable as a reflex. For more information see BCRFX / BCR/ABL1 Qualitative Diagnostic Assay with Reflex to BCR/ABL1 p190 Quantitative Assay or BCR/ABL1 p210 Quantitative Assay, Varies.

Interpretation

An interpretive report will be provided.

Method Description

The assay is performed using an automated platform, GeneXpert (Cepheid). All subsequent reactions are performed within the cartridge and the results are processed and calculated by the instrument. Quantitative, reverse transcription polymerase chain reaction (PCR) is performed with a nested PCR reaction Lot-to-lot variation in the cartridges is corrected using a calibration calculation to reference standard curve data to the IS provided by the manufacturer.(Unpublished Mayo method)

Day(s) Performed

Monday through Friday

Report Available

7 to 10 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

81206