Sign in →

Test Code MISC2MAYOMHBRP Metamycoplasma hominis, Molecular Detection, PCR, Blood

Secondary ID

65129

Useful For

Rapid, sensitive, and specific identification of Metamycoplasma hominis from whole blood

 

This test is not intended for medicolegal use.

Method Name

Real-Time Polymerase Chain Reaction (PCR) using LightCycler and Fluorescent Resonance Energy Transfer (FRET)

Reporting Name

Mycoplasma hominis PCR, B

Specimen Type

Whole Blood EDTA


Shipping Instructions


Ship specimen refrigerated.



Specimen Required


The high sensitivity of amplification by polymerase chain reaction requires the specimen to be processed in an environment in which contamination of the specimen by Metamycoplasma hominis DNA is unlikely.

 

Container/Tube:

Preferred: Lavender top (EDTA)

Acceptable: Royal blue top (EDTA), pink top (EDTA), or sterile vial containing EDTA-derived aliquot

Specimen Volume: 1 mL

Collection Instructions: Send whole blood specimen in original tube (preferred). Do not aliquot.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Whole Blood EDTA Refrigerated (preferred) 7 days
  Frozen  7 days

Reject Due To

  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Clinical Information

Metamycoplasma hominis, previously Mycoplasma hominis, has been associated with a number of clinically significant infections, although it is also part of the normal urogenital microbiota.

 

M hominis may be found in the respiratory specimens and spinal fluid of neonates. Although the clinical significance of such findings is often unclear, as spontaneous clinical recovery may occur without specific treatment. In premature infants, clinical manifestations of meningoencephalitis have been reported.

 

M hominis may play a role in some cases of pelvic inflammatory disease, usually in combination with other organisms. M hominis may be isolated from amniotic fluid of women with preterm labor, premature rupture of membranes, spontaneous term labor, or chorioamnionitis; there is evidence that it may be involved in postpartum fever or fever following abortion, usually as a complication of endometritis.

 

M hominis has rarely been associated with septic arthritis (including prosthetic joint infection), pyelonephritis, intraabdominal infection, wound infection, endocarditis, central nervous system infection (including meningoencephalitis, brain abscess, central nervous system shunt infection and subdural empyema), pneumonia, and infected pleural and pericardial effusions. Extragenital infection typically occurs in those with hypogammaglobulinemia or depressed cell-mediated immunity; in lung transplant recipients in particular, M hominis has been associated with pleuritis and mediastinitis. Recent evidence implicates donor transmission in some cases of M hominis infection in lung transplant recipients.

 

Polymerase chain reaction (PCR) detection of M hominis is sensitive, specific, and provides same-day results. Although this organism can occasionally be detected in routine plate cultures, this is neither a rapid nor a sensitive approach to detection. Specialized cultures are more time consuming than the described PCR assay. The described PCR assay has replaced conventional culture for M hominis at Mayo Clinic Laboratories due to its speed and equivalent performance to culture.

Reference Values

Not applicable

Interpretation

A positive polymerase chain reaction (PCR) result for the presence of a specific sequence found within the Metamycoplasma hominis tuf gene indicates the presence of M hominis DNA in the specimen.

 

A negative PCR result indicates the absence of detectable M hominis DNA in the specimen, but it does not rule-out infection as falsely negative results may occur due to inhibition of PCR, sequence variability underlying the primers and probes, or the presence of M hominis in quantities less than the limit of detection of the assay.

Method Description

This polymerase chain reaction (PCR) method employs a target-specific detection system including primers, as well as fluorescent resonance energy transfer (FRET) hybridization probes designed for the tuf gene of Metamycoplasma hominis. The LightCycler instrument amplifies and monitors target nucleic acid sequences by fluorescence during PCR cycling. This is an automated PCR system that can rapidly detect amplified product development. The detection of amplified products is based on the FRET principle. For FRET product detection, a hybridization probe with a donor fluorophore, fluorescein, on the 3' end is excited by an external light source, which emits light that is absorbed by a second hybridization probe with an acceptor fluorophore, LC-Red 640, on the 5' end. The acceptor fluorophore then emits light of a different wavelength that is measured with a signal that is proportional to the amount of specific PCR product. The process is completed in a closed-tube system.(Cunningham SA, Mandrekar JN, Rosenblatt JE, Patel R: Rapid PCR detection of Mycoplasma hominis, Ureaplasma urealyticum, and Ureaplasma parvum. Int J Bacteriol. 2013;2013:168742. doi: 10.1155/2013/168742)

Day(s) Performed

Monday through Friday

Report Available

3 to 4 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

87798

LOINC Code Information

Test ID Test Order Name Order LOINC Value
MHBRP Mycoplasma hominis PCR, B 42621-3

 

Result ID Test Result Name Result LOINC Value
MBSRC Specimen Source 31208-2
44131 Mycoplasma hominis PCR, B 42621-3